Schedule 12

Maximum permissible error—legal measuring instruments

In force

Schedule 12 Maximum permissible error—legal measuring instruments

(regulation 3, definition of maximum permissible error)

When comparing the error of an evidential breath analyser with the corresponding maximum permissible error, the maximum permissible error must be rounded to the value of the verification scale interval. The error of a reading indicated by an evidential breath analyser may be greater than the maximum permissible error by one least significant digit.

Table 1 Evidential breath analysers

Item

Mass concentration

Maximum permissible error for certification

Maximum permissible error for recertification

1

Not more than 0.08 g/210 litres

±0.004 g/210 litres

±0.006 g/210 litres

2

More than 0.08 g/210 litres but not more than 0.4 g/210 litres inclusive

±5% of measured concentration

±8% of measured concentration

3

More than 0.4 g/210 litres

±20% of measured concentration

±30% of measured concentration